Aerosol Valve Stem Pulling-off Strength Testing

Aerosol valve stem pulling-off strength testing is very important to aerosol valve quality. Aerosol valve has inner stem and outer stem. Labthink PARAM XLW(PC)Auto Tensile Tester can test the aerosol valve stem pulling-off force. Put the stem in between the upper and lower clamps, and pre-set the tensile speed at 50mm/min. Then, choose tensile resistance testing and start. The strength for aerosol valve to leave the valve body is the pulling-off strength. According to requirements, the final testing results should be statistical values of the maximum value of more than 3 specimens. The inner stem should be no less than 49N, and the outer stem should be no less than 40N.

Further details please feel free to contact Shirley via:
TEL: 0086-531-85061080
FAX: 0086-531-85062108
EMAIL: trade2@labthink.cn
SKYPE: shirley-labthink
MSN: shirleywu_labthink@hotmail.com

More about Labthink(http://www.labthink.cn/en):
Labthink focusing on the laboratory testing instruments R&D and production and laboratory IT services, is dedicated to provide professional quality control solutions for customers in packaging, food, pharmaceutical, medical, daily chemistry, printing, adhesive, automotive, petrochemistry, bioscience, construction, new energy and other industries.